Shopping Cart

No products in the cart.

BS EN IEC 62209-3:2019

$215.11

Measurement procedure for the assessment of specific absorption rate of human exposure to radio frequency fields from hand-held and body-mounted wireless communication devices – Vector measurement-based systems (Frequency range of 600 MHz to 6 GHz)

Published By Publication Date Number of Pages
BSI 2019 144
Guaranteed Safe Checkout
Categories: ,

If you have any questions, feel free to reach out to our online customer service team by clicking on the bottom right corner. We’re here to assist you 24/7.
Email:[email protected]

IEC 62209-3: 2019 specifies measurement protocols and test procedures for the reproducible measurement of peak spatial-average specific absorption rate (psSAR) induced inside a simplified model of a human head or body by radio-frequency (RF) transmitting devices, with a specified measurement uncertainty. Requirements are provided for psSAR assessment using vector measurement-based systems. Such systems determine the psSAR by three-dimensional (3D) field reconstruction within the volume of interest in accordance with the requirements herein for the measurement system, calibration, uncertainty assessment and validation methods. The protocols and procedures apply for the psSAR assessments covering a significant majority of people including children during use of wireless communication devices operated in close proximity to the head or body. This document is applicable to wireless communication devices intended to be used at a position near the human head or body at distances up to and including 200 mm. This document may be employed to evaluate SAR compliance of different types of wireless communication devices used next to the ear, in front of the face, mounted on the body, combined with other RF-transmitting or non-transmitting devices or accessories (e.g. belt-clip), or embedded in garments. The overall applicable frequency range is from 600 MHz to 6 GHz. The system validation procedures provided within this document cover frequencies from 600 MHz to 6 GHz. With a vector measurement-based system this document can be employed to evaluate SAR compliance of different types of wireless communication devices. The wireless communication device categories covered include but are not limited to mobile telephones, cordless microphones, auxiliary broadcast devices and radio transmitters in personal computers, desktop and laptop devices, multi-band, multi-antenna, and push-to-talk devices. Key Words: Human Exposure, Hand-Held and Body Mounted Wireless Communication Devices.

PDF Catalog

PDF Pages PDF Title
2 undefined
5 Annex ZA(normative)Normative references to international publicationswith their corresponding European publications
7 English
CONTENTS
14 FOREWORD
16 INTRODUCTION
17 1 Scope
2 Normative references
18 3 Terms and definitions
19 4 Symbols and abbreviated terms
5 Overview of the measurement procedure
20 Figures
Figure 1 – Evaluation plan checklist
21 Tables
Table 1 – Evaluation plan checklist
22 6 Measurement system specifications
6.1 General requirements
24 6.2 Phantom specifications
6.2.1 Head phantom specifications – shell
6.2.2 Body phantom specifications – shell
6.2.3 Tissue-equivalent medium properties
6.3 Measurement system requirements
6.3.1 General
6.3.2 Scanning measurement system specifications
25 6.3.3 Array measurement system specifications
26 6.4 Device holder specification
27 6.5 Reconstruction algorithm and peak spatial-averaging specifications
7 Protocol for SAR assessments
7.1 Measurement preparation
7.1.1 General
7.1.2 Preparation of tissue-equivalent medium
Figure 2 – Illustration of the shape and orientation relative to a curved phantomsurface of the distorted cubic volume for computing psSAR
28 7.1.4 Preparation of the device under test (DUT)
7.1.5 Operating modes
7.1.6 Position of the DUT in relation to the phantom
7.1.7 Positions of the DUT in relation to the flat phantom for large DUT
29 7.1.8 Test frequencies for DUT
7.2 Tests to be performed
Figure 3 – Measurements performed by shifting a large device over the efficientmeasurement area of the system including overlapping areas –in this case: six tests performed
30 7.3 General measurement procedure
7.3.1 General
7.3.2 Measurement procedure for scanning systems
31 7.3.3 Measurement procedure for array systems
7.4 SAR measurements for simultaneous transmission
7.4.1 General
32 7.4.2 SAR measurements for uncorrelated signals
Figure 4 – Flow chart for SAR measurements of uncorrelated signals at different frequencies using a measurement system able to distinguish between different frequency components (Method 2)
33 Figure 5 – Illustration of the amplitude spectrum, as function of frequency, for simultaneously transmitted signals of multiple frequency bands emitted by a DUT
34 Figure 6 – Illustration of a completely covered signal bandwidth Bs by the measurement system analysis bandwidth Ba at single transmission mode
Figure 7 – Illustration of a completely covered signal bandwidthsBsi (for i = 2 to N) by the measurement system analysis bandwidth Bafor simultaneous multiple-frequency transmission mode
Figure 8 – Illustration of a non-coverage of the signal bandwidthsBsi (for i = 2 to N) by the measurement system analysis bandwidth Bafor simultaneous multiple-frequency transmission mode
35 Figure 9 – Illustration of a partial-coverage of the signal bandwidthsBsi (for i = 2 to N) by the measurement system analysis bandwidth Bafor simultaneous multiple-frequency transmission mode
Figure 10 – Illustration of reduction of the measurement system analysisbandwidth Ba to cover only one signal bandwidth Bsi (for i = 1 to N)for simultaneous multiple-frequency transmission mode
Figure 11 – Illustration of increasing or moving the measurement systemanalysis bandwidth Ba to cover one or more signal bandwidth Bsi (for i = 1 to N)for simultaneous multiple-frequency transmission mode
36 7.4.3 SAR measurements for correlated signals
37 8 Measurement uncertainty estimation
8.1 General
8.2 Requirements on the measurement uncertainty evaluation
38 8.3 Description of measurement uncertainty models
8.3.1 General
39 8.3.2 Uncertainty models for array measurement system and scanning measurement systems
40 8.3.3 Example uncertainty budget templates
41 Table 2 – Uncertainty budget template for the evaluation of the measurement system uncertainty of the 1 g or 10 g psSAR to be carried out by the system manufacturer
42 Table 3 – Uncertainty budget template for evaluating the uncertaintyin the measured value of 1 g SAR or 10 g SAR from a DUT
43 Table 4 – Uncertainty budget template for evaluating the uncertaintyin the measured value of 1 g SAR or 10 g SAR from a validation antenna
44 9 Measurement report
Table 5 – Uncertainty budget template for evaluating the uncertainty inthe measured value of 1 g SAR or 10 g SAR from the system check
45 Annex A (normative)Phantom specifications
A.1 SAM phantom specifications
A.1.1 Justification
A.1.2 SAM phantom geometry
A.1.3 Tissue-equivalent medium
46 A.2 Flat phantom specifications
Figure A.1 – Sagittally-bisected phantom with extended perimeter,used for scanning measurement systems
47 A.3 Specific phantoms
Figure A.2 – Dimensions of the elliptical phantom
48 A.4 Tissue-equivalent medium
Table A.1 – Dielectric properties of the tissue-equivalent medium
49 Annex B (normative)Calibration and characterization of dosimetric probes
B.1 General
B.2 Types of calibration
B.2.1 Amplitude calibration with analytical fields
50 B.2.2 Amplitude and phase calibration by transfer calibration
Table B.1 – Uncertainty analysis for single-probe calibration in waveguide
51 Table B.2 – Uncertainty analysis for transfer calibration of array systems
52 B.2.3 Amplitude and phase calibration using numerical reference
53 Table B.3 – Uncertainty analysis of transfer calibration of array systems
54 Annex C (informative)Field reconstruction techniques
C.1 General
C.2 Objective of field reconstruction techniques
C.3 Background
55 Figure C.1 – Coordinate system for 2D planar measurement-system
Figure C.2 – Generic configuration of SAR measurement system
56 C.4 Reconstruction techniques
C.4.1 Expansion techniques
57 C.4.2 Source reconstruction techniques
C.4.3 Source base function decomposition
C.4.4 Phase reconstruction
Figure C.3 – Schematic representation of 2D planar measurement-based SARsystem and its coordinate system
58 C.5 Source reconstruction and SAR estimation from fields measured outside the phantom
C.6 Additional considerations for field reconstruction in scanning systems
Figure C.4 – Source reconstruction from fields outside a phantom
59 Annex D (normative)SAR measurement system verification and system validation
D.1 Objectives and purpose
D.1.1 General
D.1.2 Objectives and purpose of system check
D.1.3 Objectives of system validation
60 D.2 SAR measurement setup and procedure for system check and system validation
D.2.1 General
D.2.2 Power measurement setups
61 Figure D.1 – Recommended power measurement setupfor system check and system validation
62 D.2.3 Procedure to measure and normalize SAR
Figure D.2 – Equipment setup for measurement of forwardpower Pf and forward coupled power Pfc
63 Figure D.3 – Equipment setup for measuring the shorted reverse coupled power Prcs
Figure D.4 – Equipment setup for measuring the powerwith the reference antenna connected
64 D.2.4 Power measurement uncertainty
65 Figure D.5 – Port numbering for the S-parameter measurementsof the directional coupler
Table D.1 – Example of power measurement uncertainty in %
66 D.3 System check
D.3.1 System check antennas and test conditions
D.3.2 System check antennas and test conditions for scanning systems
D.3.3 System check antennas and test conditions for array systems
67 D.3.4 System check acceptance criteria
D.4 System validation
D.4.1 Validation of array systems and scanning systems
D.4.2 Requirements for system validation antennas and test conditions
D.4.3 Requirements for array systems and scanning systems
69 D.4.4 Test positions for system validation
Table D.2 – Modulations and multiplexing modes used by radio systems
70 Figure D.6 – SAM masks for positioning dipole antennas and VPIFAs on the head phantoms, including holes where the antenna spacer is inserted
71 Figure D.7 – Flat masks for positioning VPIFAs on the flat phantoms, including a hole in the centre where the VPIFA spacer is inserted
72 Figure D.8 – Dipole showing the distance of s = 15 mm
Figure D.9 – 2-PEAK CPIFA showing the fixed distance of s = 7 mm
73 Figure D.10 – VPIFA positioned showing the fixed distance of s = 2 mm
74 Figure D.11 – System check and validation locations for the flat phantom
75 Figure D.12 – System check and validation locations for the head phantom
76 D.4.5 System validation procedure based on peak spatial-average SAR
Figure D.13 – Definition of rotation angles for dipoles
77 Table D.3 – Peak spatial-average SAR (psSAR) averaged over 1 g and 10 g values for the flat phantom filled with tissue-equivalent medium for the antennas specified in Annex F
78 Table D.4 – Peak spatial-average SAR (psSAR) averaged over 1 g and 10 g values for antenna generating two peaks on the flat phantom filled with tissue-equivalent medium for the antennas specified in Annex F
79 Table D.5 – Peak spatial-average SAR (psSAR) averaged over 1 g and 10 g values on the head leftand right phantom for the antennas specified in Annex F
84 D.4.6 On-site system validation after installation
Table D.6 – Peak spatial-average SAR (psSAR) averaged over 1 g and 10 g values for antenna generating two peaks on the head left and right phantom for the antennas specified in Annex F. Modulations are as specified in Table D.2
Table D.7 – Set of randomised tests for on-site system validation using flat phantom 1 g and 10 g psSAR, normalized to 1 W forward power, using the antennas specified in Annex F
85 D.4.7 System validation acceptance criteria
Table D.8 – Set of tests for on-site system validation using left and right head phantoms for 1 g and 10 g psSAR for the antennas specified in Annex F
87 Annex E (informative)Interlaboratory comparisons
E.1 Purpose
E.2 Monitor laboratory
E.3 Phantom set-up
E.4 Reference devices
88 E.5 Power set-up
E.6 Interlaboratory comparison – Procedure
89 Annex F (normative)System validation antennas
F.1 General requirements
F.2 Return loss requirements
90 F.3 Standard dipole antenna
Table F.1 – Return loss values for antennas specified in Annex Fand flat phantom filled with tissue-equivalent medium
91 Table F.2 – Mechanical dimensions of the reference dipoles
92 Figure F.1 – Mechanical details of the standard dipole
93 F.4 VPIFA
94 Figure F.2 – VPIFA validation antenna
95 F.5 2-PEAK CPIFA
Table F.3 – Dimensions for VPIFA antennas at different frequencies
Table F.4 – Dielectric properties of the dielectric layers for VPIFA antennas
97 Figure F.3 – 2-PEAK CPIFA at 2 450 MHz
98 Figure F.4 – Detail of the tuning structure and matching structure
Table F.5 – Thickness of substrates and planar metallization
Table F.6 – Dielectric properties of FR4
99 F.6 Additional antennas
Table F.7 – Values for the antenna dimensions in Figures F.4 and F.5
100 Annex G (normative)SAR calibration of reference antennas
G.1 Purpose
Figure G.1 – Measurement setup for waveguide calibration of dosimetric probe,and similar setup (same tissue-equivalent liquid, dielectric spacer,power sensors and coupler) for antenna calibration
101 G.2 Parameters or quantities and ranges to be determined by calibration method
G.3 Reference antenna calibration setup
Figure G.2 – Setup for calibration of a reference antenna
102 G.4 Reference antenna calibration procedure
G.4.1 Verification of return loss
G.4.2 Calibration of reference antennas: step-by-step procedure
103 G.4.3 Uncertainty budget of reference antenna calibration
104 Table G.1 – Example uncertainty budget for reference dipole antennacalibration for 1 g and 10 g averaged SAR (750 MHz to 3 GHz)
105 Table G.2 – Example uncertainty budget for reference antenna calibration (PIFA)for 1 g and 10 g averaged SAR (750 MHz to 3 GHz)
106 Table G.3 – Example uncertainty budget for reference antenna (dipole) calibrationfor 1 g and 10 g averaged SAR (3 GHz to 6 GHz)
107 G.5 Method and uncertainties for the transfer of calibration between two of more antennas of the same type using the array system
108 Figure G.3 – Method for the transfer of calibration between two antennasof the same type using the array system
109 Table G.4 – Example uncertainty budget for the calibration ofan antenna using the transfer method, as percentages
110 Annex H (informative)General considerations on uncertainty estimation
H.1 Concept of uncertainty estimation
111 H.2 Type A and Type B evaluations
H.3 Degrees of freedom and coverage factor
112 H.4 Combined and expanded uncertainties
113 H.5 Analytical reference functions
114 Table H.1 – Parameters of analytical reference functionsand associated reference peak 10 g SAR value
116 Annex I (normative)Evaluation of measurement uncertainty of SAR results from scanning vector measurement-based systems with single probes
I.1 Measurement uncertainties to be evaluated by the system manufacturer MM
I.1.1 General
I.1.2 Calibration CF
I.1.3 Isotropy ISO
117 I.1.4 System linearity LIN
I.1.5 Sensitivity limit SL
I.1.6 Boundary effect BE
118 I.1.7 Readout electronics RE
I.1.8 Response time RT
I.1.9 Probe positioning PP
I.1.10 Sampling error SE
119 I.1.11 Phantom shell PS
I.1.12 Tissue-equivalent medium parameters MAT
121 I.1.13 Measurement system immunity/secondary reception MSI
I.2 Uncertainty of reconstruction corrections and post-processing to be specified by the manufacturer MN
I.2.1 General
I.2.2 Evaluation of uncertainty due to reconstruction REC
122 I.2.3 Impact of noise on reconstruction POL
I.2.4 SAR averaging SAV
I.2.5 SAR scaling SARS
123 I.2.6 SAR correction for deviations in permittivity and conductivity SC
124 I.3 Uncertainties that are dependent on the DUT MD
I.3.1 General
I.3.2 Probe coupling with the DUT PC
I.3.3 Modulation Response MOD
125 I.3.4 Integration time IT
I.3.5 Measured SAR drift SD
I.4 Uncertainties related to the measurement environment ME
I.4.1 General
I.4.2 Device holder DH
126 I.4.3 Device positioning DP
I.4.4 RF ambient conditions AC
I.4.5 Measurement system drift and noise DN
127 I.5 Uncertainties of validation antennas MV
I.5.1 General
I.5.2 Deviation of experimental antennas DEX
I.5.3 Power measurement uncertainty PMU
I.5.4 Other uncertainty contributions when using validation antennas OVS
Figure I.1 – Illustration of SAR measurement results during 8 hand the centred moving average
128 Annex J (normative)Evaluation of the measurement system uncertainty of fixed arrayor scanning array vector measurement-based systems
J.1 Measuring system uncertainties to be evaluated by the manufacturer MM
J.1.1 General
J.1.2 Calibration CF
J.1.3 Isotropy ISO
129 J.1.4 Mutual sensor coupling MSC
130 J.1.5 Scattering due to the presence of the array AS
131 J.1.6 System linearity LIN
J.1.7 Sensitivity limit SL
J.1.8 Boundary effect BE
132 J.1.9 Readout electronics RE
J.1.10 Response time RT
J.1.11 Probe position PP
133 J.1.12 Sampling error SE
J.1.13 Array boundaries AB
134 J.1.14 Phantom shell PS
J.1.15 Tissue-equivalent medium parameters MAT
136 J.1.16 Phantom homogeneity HOM
137 J.1.17 Measurement system immunity/secondary reception MSI
J.2 Uncertainty of reconstruction, corrections, and post-processing to be specified by the manufacturer MN
J.2.1 General
J.2.2 Evaluation of uncertainty due to reconstruction REC
J.2.3 Impact of noise on reconstruction POL
J.2.4 SAR averaging SAV
J.2.5 SAR scaling SARS
J.2.6 SAR correction for deviations in permittivity and conductivity SC
J.3 Measurement system uncertainties that are dependent on the DUT MD
J.3.1 General
J.3.2 Probe or probe-array coupling with the DUT PC
138 J.3.3 Modulation response MOD
J.3.4 Integration time IT
J.3.5 Measurement system drift and noise DN
J.4 Uncertainties related to the source or noise ME
J.4.1 General
J.4.2 Device holder DH
J.4.3 Device positioning DP
139 J.4.4 RF ambient conditions AC
J.4.5 Measurement system drift and noise DN
J.5 Uncertainties of validation antennas MV
J.5.1 General
J.5.2 Deviation of experimental antennas DEX
J.5.3 Power measurement uncertainty PMU
J.5.4 Other uncertainty contributions when using validation antennas OVS
140 Bibliography
BS EN IEC 62209-3:2019
$215.11