Shopping Cart

No products in the cart.

BS EN ISO 22232-1:2020 – TC

$246.62

Tracked Changes. Non-destructive testing. Characterization and verification of ultrasonic test equipment – Instruments

Published By Publication Date Number of Pages
BSI 2020 122
Guaranteed Safe Checkout
Category:

If you have any questions, feel free to reach out to our online customer service team by clicking on the bottom right corner. We’re here to assist you 24/7.
Email:[email protected]

This document specifies methods and acceptance criteria within the frequency range of 0,5 MHz to 15 MHz, for assessing the electrical performance of digital ultrasonic instruments for pulse operation using A-scan display, for manual ultrasonic non-destructive testing with single- or dual-transducer probes. This document is also applicable for multi-channel instruments. This document can partly be applicable to ultrasonic instruments in automated systems, but other tests can be needed to ensure satisfactory performance.

This document excludes ultrasonic instruments for continuous waves.

This document also excludes ultrasonic phased array instruments, see e.g. ISO 18563-1. If a phased array instrument has dedicated connectors for single- or dual-transducer probes this document is applicable for these channels.

PDF Catalog

PDF Pages PDF Title
3 National foreword
Compliance with a British Standard cannot confer immunity from legal obligations.
5 ForewordEuropean foreword
Endorsement notice
8 Foreword
9 Non-destructive testing — Characterization and
Instruments
1 Scope
2 Normative references
3 Terms and definitions
10 3.1
3.2
3.4
calibrated dB-switch
3.53.3
3.4
3.63.5
3.7
dynamic range
3.83.6
11 3.93.7
3.103.8
time it takes the proportional gate output to fall from 90 % to 10 % of its peak value
3.11
3.123.9
3.133.10
time for which the proportionalanalogue output (3.1) is above 90 50 % of its peakmaximum output following a signal in the monitor gate
3.143.11
3.15
3.16
3.173.12
3.18
3.19
3.20
12 3.22
pulse duration
3.23
pulse repetition frequency
3.24
pulse rise time
3.25
pulse reverberation
3.263.13
3.273.14
3.28
rise time of proportional output
3.293.15
3.30
time-dependent gain TDG
3.31
short pulse
3.32
suppression
3.333.16
13 4 Symbols
Table 1 — Symbols
14 6 Manufacturer’s technical specification for ultrasonic instruments
19 7 Performance requirements for ultrasonic instruments
23 8 Group 1 tests
8.1 Equipment required for group 1 tests
24 8.2 Stability against temperature
8.2 Battery operational time
8.2.1 Procedure
25 Figure 1 — Set up for measuring stability against temperature
26 8.2.2 Acceptance criterion
8.3 Stability after warm‑up time
8.3.1 Procedure
8.3.2 Acceptance criteria
8.4 Stability against temperature
8.4.1 Procedure
27 Figure 1 — Setup for measuring stability against temperature
28 Figure 2 — Circuit to protect the instrument from the transmitter pulse
8.2.28.4.2 Acceptance criterion
8.38.5 Stability against voltage variation
8.3.18.5.1 Procedure
29 8.3.28.5.2 Acceptance criterion
8.6 Time base deviation
8.6.1 Procedure
30 Figure 3 — Setup of equipment for multiple tests
31 8.6.2 Acceptance criterion
8.48.7 Transmitter pulse parameters
8.4.18.7.1 General
8.4.28.7.2 Pulse repetition frequency
8.4.2.28.7.2.2 Acceptance criterion
8.4.38.7.3 Effective output impedance
8.4.3.18.7.3.1 Procedure
32 8.4.3.28.7.3.2 Acceptance criterion
8.4.4 Transmitter pulse frequency spectrum
8.4.4.1 Procedure
8.4.4.2 Acceptance criterion
8.58.8 Receiver
8.5.18.8.1 General
8.5.28.8.2 Cross talk from transmitter to receiver during transmission
33 Figure 4 — Equipment setup used to measure cross talk
8.5.2.28.8.2.2 Acceptance criterion
8.5.38.8.3 Dead time after transmitter pulse
34 Figure 5 — Equipment setup used to measure dead time after the transmitter pulse
35 8.5.3.2 Acceptance criterion
Figure 6 — Waveform used to measure dead time after the transmitter pulse as seen on the instrument screen during the test
8.8.3.2 Acceptance criterion
8.5.48.8.4 Dynamic range and maximum input voltage
36 Figure 7 — Test waveform generated by general purpose equipment setup
37 8.5.4.28.8.4.2 Acceptance criteria
8.5.58.8.5 Receiver input impedance
8.5.5.18.8.5.1 Procedure
8.5.5.28.8.5.2 Acceptance criterion
8.5.68.8.6 Time-dependent‑corrected gain (TDGTCG)
38 8.5.6.28.8.6.2 Acceptance criterion
8.5.7 Temporal resolution
8.5.7.1 Procedure
The widest band setting of the equipment is selected. Set the equipment in Figure 6 to generate two single cycle measurement pulses with centre frequency fo measured in 9.5.2 for the frequency band chosen. These pulses should follow each other at a di…
39 Increase the amplitude of the first measurement pulse by 20 dB, while maintaining the amplitude of the second pulse as 80 % of screen height. Decrease the distance between the two measurement pulses until the dip between both of them is 6 dB (relative…
8.9 Gates
8.9.1 General
Figure 8 — Timing diagram of signals used to test a monitor gate
40 8.9.2 Gates with value output
8.9.2.1 Linearity of the amplitude in the gate
Table 3 — Expected gate amplitude for specified attenuator settings
8.9.2.1.2 Acceptance criterion
8.9.2.2 Linearity of time of flight in the gate
41 Table 4 — Expected gate TOF for specified positions in the screen width
8.9.3 Gates with analogue output
8.9.3.1.1 Procedure
8.9.3.1.2 Acceptance criterion
42 8.9.3.2 Linearity of analogue output
Table 5 — Expected output voltage for specified attenuator settings
8.9.3.2.2 Acceptance criterion
8.9.3.3 Influence of the signal position within the gate
8.9.3.3.1 Procedure
43 8.9.3.3.2 Acceptance criterion
8.9.3.4 Rise time, fall time, delay time and hold time of analogue output
8.9.3.4.1 Procedure
8.5.7.28.9.3.4.2 Acceptance criterion
8.9.4 Gates with alarm output
44 8.6.28.9.4.2 Response threshold and switching hysteresis with a fixed monitor threshold
8.9.4.3 Delay time and hold time of the gate alarm
8.6.3 Hold time of the switched output
8.6.3.28.9.4.3.2 Acceptance criterioncriteria
8.7.1 Impedance of proportional output
8.7.1.1 Procedure
45 8.7.2 Linearity of proportional output
8.7.2.1 Procedure
Table 3 — Expected output voltage for specified attenuator settings
46 8.7.2.2 Acceptance criterion
8.7.3 Frequency response of proportional gate output
8.7.3.1 Procedure
8.7.3.2 Acceptance criterion
8.7.4 Noise on proportional gate output
8.7.4.1 Procedure
8.7.4.2 Acceptance criterion
8.7.5 Influence of the measurement signal position within the gate
8.7.5.1 Procedure
47 8.7.5.2 Acceptance criterion
8.7.6 Effect of pulse shape on the proportional gate output
8.7.6.1 Procedure
8.7.6.2 Acceptance criterion
8.7.7 Rise, fall and hold time of proportional gate output
8.7.7.1 Procedure
8.7.7.2 Acceptance criterion
48 8.8.1 General
8.8.2 Linearity of time-base for digital ultrasonic instruments
8.8.2.1 Procedure
8.8.2.2 Acceptance criterion
8.8.3 Digitisation sampling error
8.8.3.1 Procedure
8.10.1 Procedure
8.10.1.1 General
49 8.10.1.2 Method A
8.10.1.3 Method B
Figure 9 — Waveform used with method B to measure the digitisation sampling error
50 8.8.2.28.10.2 Acceptance criterion
8.8.48.11 Response time of digital ultrasonic instruments
8.8.4.18.11.1 ProcedureGeneral
8.11.2 Procedure
51 Figure 10 — Timing diagram showing how to measure the response time of digital ultrasonic instruments
8.8.4.28.11.3 Acceptance criterion
9 Group 2 tests
9.1 Equipment required for group 2 tests
52 9.2 Physical state and external aspects
9.2.1 Procedure
9.2.2 Acceptance criterion
9.3.1 General
9.3.2 Stability after warm-up time
9.3.2.1 Procedure
9.3.2.2 Acceptance criteria
53 9.3.3 Display jitter
9.3.3.1 Procedure
9.3.3.2 Acceptance criteria
9.3.4 Stability against voltage variations
9.3.4.1 Procedure
9.3.4.2 Acceptance criteria
9.4.1 General
9.4.29.3 Transmitter voltage, pulse rise time, reverberation and duration
9.4.2.19.3.1 Procedure
54 Figure 11 — Instrumentation for pulse measurements
55 Figure 12 — Transmitter pulse parameters to be measured
56 9.4.2.29.3.2 Acceptance criteria
9.59.4 Receiver
9.5.19.4.1 General
9.5.29.4.2 Amplifier frequencyFrequency response
57 Figure 13 — Receiver section frequency characteristics
9.5.2.2 Acceptance criteria
58 9.4.2.2 Acceptance criteria
9.5.3 Equivalent input noise
9.5.3.1 Procedure
9.4.3 Noise level
9.4.3.1 Procedure
9.4.3.2 Method A
9.4.3.3 Method B
59 9.5.3.29.4.3.4 Acceptance criterion
9.5.4 Accuracy of calibrated attenuator
9.4.4 Gain linearity
60 9.5.4.2 Acceptance criteria
9.4.4.2 Acceptance criteria
9.5.5 Linearity of vertical display
9.4.5 Vertical display linearity
61 Table 4Table 6 — Acceptance levels for vertical display linearity
9.6.1 Procedure
9.6.2 Acceptance criterion
62 Key
Figure 2 — Circuit to protect equipment from the transmitter pulse
Figure 3 — Equipment set-up used to measure cross-talk damping
63 Figure 4 — Equipment set-up used to measure dead time after the transmitter pulse
64 Key
Figure 5 — Waveform used to measure dead time after the transmitter pulse as seen on the instrument screen during the test
65 Figure 6 — General purpose set-up for equipment
66 Key
Figure 7 — Test waveform generated by general purpose equipment set-up
Figure 8 — Set-up of equipment for tests on monitor gate
67 Figure 9 — Timing diagram of signals used to test monitor gate
68 Key
Figure 10 — Timing diagram showing how to measure the response time of digital flaw detectors
69 Figure 11 — Transmitter pulse parameters to be measured
70 Key
Figure 12 — Signals used to test time base linearity
71 Special conditions for ultrasonic instruments with logarithmic amplifiers
A.1 IntroductionGeneral
A.2 Basic requirements
A.2.1 Measuring accuracy
A.2.2 Vertical display “linearity
A.3A.2.3 Tests
72 Bibliography
74 National foreword
76 European foreword
79 Foreword
81 1 Scope
2 Normative references
3 Terms and definitions
83 4 Symbols
84 5 General requirements of conformity
6 Manufacturer’s technical specification for ultrasonic instruments
87 7 Performance requirements for ultrasonic instruments
89 8 Group 1 tests
8.1 Equipment required for group 1 tests
90 8.2 Battery operational time
8.2.1 Procedure
8.2.2 Acceptance criterion
8.3 Stability after warm-up time
8.3.1 Procedure
8.3.2 Acceptance criteria
91 8.4 Stability against temperature
8.4.1 Procedure
93 8.4.2 Acceptance criterion
8.5 Stability against voltage variation
8.5.1 Procedure
94 8.5.2 Acceptance criterion
8.6 Time base deviation
8.6.1 Procedure
96 8.6.2 Acceptance criterion
8.7 Transmitter pulse parameters
8.7.1 General
8.7.2 Pulse repetition frequency
8.7.3 Effective output impedance
97 8.8 Receiver
8.8.1 General
8.8.2 Cross talk from transmitter to receiver during transmission
98 8.8.3 Dead time after transmitter pulse
100 8.8.4 Dynamic range and maximum input voltage
101 8.8.5 Receiver input impedance
102 8.8.6 Time-corrected gain (TCG)
103 8.9 Gates
8.9.1 General
104 8.9.2 Gates with value output
106 8.9.3 Gates with analogue output
108 8.9.4 Gates with alarm output
109 8.10 Highest digitized frequency
8.10.1 Procedure
110 8.10.2 Acceptance criterion
8.11 Response time of digital ultrasonic instruments
8.11.1 General
8.11.2 Procedure
111 8.11.3 Acceptance criterion
9 Group 2 tests
9.1 Equipment required for group 2 tests
112 9.2 Physical state and external aspects
9.2.1 Procedure
9.2.2 Acceptance criterion
9.3 Transmitter voltage, pulse rise time and duration
9.3.1 Procedure
115 9.3.2 Acceptance criteria
9.4 Receiver
9.4.1 General
9.4.2 Frequency response
117 9.4.3 Noise level
118 9.4.4 Gain linearity
9.4.5 Vertical display linearity
120 Annex A (normative) Special conditions for ultrasonic instruments with logarithmic amplifiers
121 Bibliography
BS EN ISO 22232-1:2020 - TC
$246.62